tem:procedures [2016/02/08 15:24]
admin [Alignment]
tem:procedures [2021/10/25 08:53] (current)
Line 25: Line 25:
   * COND STIG: make round with DEF   * COND STIG: make round with DEF
   * STD FOCUS adjust Z (minimal contrast, no hallo, or no motion with IMAGE WOBB)   * STD FOCUS adjust Z (minimal contrast, no hallo, or no motion with IMAGE WOBB)
-  * Tilt and shift purity+  * Tilt and shift purity (at correct focus, and diffraction settings!)
   * OBJ wobler (F4) make feature stationary with DEF(BRIGHT TILT)   * OBJ wobler (F4) make feature stationary with DEF(BRIGHT TILT)
   * make fft round with DEF(OBJ STIG)   * make fft round with DEF(OBJ STIG)
Line 46: Line 46:
 ==== Beam shower ==== ==== Beam shower ====
  
-  * ?+  * Camera length 150cm 
 +  * CLA out, SAA #4, HCA #4 
 +  * CL1 0x8000, CL2 0x6000 (load STEMBeamShower.flc in FLC Panel) 
 +  * wait 10-15 min
  
-==== After (brief) power failure  ====+==== After power failure  ====
  
   * reset chiller complaining about closed outlet   * reset chiller complaining about closed outlet
 +  * (on-off HT breaker at the back of power supply unit)
   * press (longer) "I" on the front of the microscope   * press (longer) "I" on the front of the microscope
 +  * (Lenses on - front of the microscope)
 +  * (EDX detector controller on-off, while SIP is not running (ACD on))
   * watch vacuum   * watch vacuum
 +  * (HT conditioning: switch knob to Cond, Auto procedure start, switch knob back to Operate)
   * HT -> Auto HT & Emission -> Start up   * HT -> Auto HT & Emission -> Start up
  