* COND STIG: make round with DEF | * COND STIG: make round with DEF | ||
* STD FOCUS adjust Z (minimal contrast, no hallo, or no motion with IMAGE WOBB) | * STD FOCUS adjust Z (minimal contrast, no hallo, or no motion with IMAGE WOBB) | ||
- | * Tilt and shift purity | + | * Tilt and shift purity |
* OBJ wobler (F4) make feature stationary with DEF(BRIGHT TILT) | * OBJ wobler (F4) make feature stationary with DEF(BRIGHT TILT) | ||
* make fft round with DEF(OBJ STIG) | * make fft round with DEF(OBJ STIG) | ||
Line 46: | Line 46: | ||
==== Beam shower ==== | ==== Beam shower ==== | ||
- | * ? | + | * Camera length 150cm |
+ | * CLA out, SAA #4, HCA #4 | ||
+ | * CL1 0x8000, CL2 0x6000 (load STEMBeamShower.flc in FLC Panel) | ||
+ | * wait 10-15 min | ||
- | ==== After (brief) | + | ==== After power failure |
* reset chiller complaining about closed outlet | * reset chiller complaining about closed outlet | ||
+ | * (on-off HT breaker at the back of power supply unit) | ||
* press (longer) " | * press (longer) " | ||
+ | * (Lenses on - front of the microscope) | ||
+ | * (EDX detector controller on-off, while SIP is not running (ACD on)) | ||
* watch vacuum | * watch vacuum | ||
+ | * (HT conditioning: | ||
* HT -> Auto HT & Emission -> Start up | * HT -> Auto HT & Emission -> Start up | ||