tem:procedures [2015/12/09 08:24]
admin [Normal vacuum levels]
tem:procedures [2021/10/25 08:53] (current)
Line 21: Line 21:
 ==== Alignment ==== ==== Alignment ====
  
-  * GUNA(F3)center with DEF @ Spot1 and SHIFT Spot5, or Anode wobler(F5)+  * GUNA(F3) Spot1 center with SHIFT <-> BRIGHT TILT Spot5 center with SHIFT 
 +  * GUNA(F3) and Anode wobler(F5) make symetric with DEF
   * COND STIG: make round with DEF   * COND STIG: make round with DEF
-  * Tilt and shift purity 
   * STD FOCUS adjust Z (minimal contrast, no hallo, or no motion with IMAGE WOBB)   * STD FOCUS adjust Z (minimal contrast, no hallo, or no motion with IMAGE WOBB)
 +  * Tilt and shift purity (at correct focus, and diffraction settings!)
   * OBJ wobler (F4) make feature stationary with DEF(BRIGHT TILT)   * OBJ wobler (F4) make feature stationary with DEF(BRIGHT TILT)
   * make fft round with DEF(OBJ STIG)   * make fft round with DEF(OBJ STIG)
Line 30: Line 31:
   * Filter->Slit->IN move slit to center using position   * Filter->Slit->IN move slit to center using position
   * back to IMAGE, center circle with Isochromacity   * back to IMAGE, center circle with Isochromacity
-  * center diffraction win DEF(PLA)+  * center diffraction with DEF(PLA) 
 +  * focus diffraction with DIFF FOCUS at Alpha 3, camera length 200cm, condensor fully CW
  
  
 ==== After bakeout ==== ==== After bakeout ====
  
 +  * remove mechanical safeties from EDX detector valve and beam valve
   * Cycle BEAM VALVE with HT OFF   * Cycle BEAM VALVE with HT OFF
   * Open EDX detector valve with BEAM VALVE closed   * Open EDX detector valve with BEAM VALVE closed
 +  * Conditioning knob -> COND
   * HT->HT Conditioning->Auto Procedure->Start (it takes ~60 min)   * HT->HT Conditioning->Auto Procedure->Start (it takes ~60 min)
 +  * Conditioning knob -> OPERATE
  
 ==== Beam shower ==== ==== Beam shower ====
  
-  * ?+  * Camera length 150cm 
 +  * CLA out, SAA #4, HCA #4 
 +  * CL1 0x8000, CL2 0x6000 (load STEMBeamShower.flc in FLC Panel) 
 +  * wait 10-15 min
  
-==== After (brief) power failure  ====+==== After power failure  ====
  
   * reset chiller complaining about closed outlet   * reset chiller complaining about closed outlet
-  * press (longer) I on the front of the microscope+  * (on-off HT breaker at the back of power supply unit) 
 +  * press (longer) "Ion the front of the microscope 
 +  * (Lenses on - front of the microscope) 
 +  * (EDX detector controller on-off, while SIP is not running (ACD on))
   * watch vacuum   * watch vacuum
 +  * (HT conditioning: switch knob to Cond, Auto procedure start, switch knob back to Operate)
   * HT -> Auto HT & Emission -> Start up   * HT -> Auto HT & Emission -> Start up
  
tem/procedures.1449649464.txt.gz · Last modified: 2021/10/25 08:53 (external edit)