This shows you the differences between two versions of the page.
Next revision | Previous revision | ||
tem:procedures [2015/11/13 23:14] 127.0.0.1 external edit |
tem:procedures [2021/10/25 08:53] (current) |
||
---|---|---|---|
Line 21: | Line 21: | ||
==== Alignment ==== | ==== Alignment ==== | ||
- | * GUNA(F3): center with DEF @ Spot1 and SHIFT @ Spot5, or Anode wobler(F5) | + | * GUNA(F3) |
+ | * GUNA(F3) and Anode wobler(F5) | ||
* COND STIG: make round with DEF | * COND STIG: make round with DEF | ||
- | * Tilt and shift purity | ||
* STD FOCUS adjust Z (minimal contrast, no hallo, or no motion with IMAGE WOBB) | * STD FOCUS adjust Z (minimal contrast, no hallo, or no motion with IMAGE WOBB) | ||
+ | * Tilt and shift purity (at correct focus, and diffraction settings!) | ||
* OBJ wobler (F4) make feature stationary with DEF(BRIGHT TILT) | * OBJ wobler (F4) make feature stationary with DEF(BRIGHT TILT) | ||
* make fft round with DEF(OBJ STIG) | * make fft round with DEF(OBJ STIG) | ||
- | * Filter-> | + | |
- | * center diffraction | + | |
+ | * back to IMAGE, | ||
+ | * center diffraction | ||
+ | * focus diffraction with DIFF FOCUS at Alpha 3, camera length 200cm, condensor fully CW | ||
==== After bakeout ==== | ==== After bakeout ==== | ||
+ | * remove mechanical safeties from EDX detector valve and beam valve | ||
* Cycle BEAM VALVE with HT OFF | * Cycle BEAM VALVE with HT OFF | ||
* Open EDX detector valve with BEAM VALVE closed | * Open EDX detector valve with BEAM VALVE closed | ||
+ | * Conditioning knob -> COND | ||
* HT->HT Conditioning-> | * HT->HT Conditioning-> | ||
+ | * Conditioning knob -> OPERATE | ||
==== Beam shower ==== | ==== Beam shower ==== | ||
- | * ? | + | * Camera length 150cm |
+ | * CLA out, SAA #4, HCA #4 | ||
+ | * CL1 0x8000, CL2 0x6000 (load STEMBeamShower.flc in FLC Panel) | ||
+ | * wait 10-15 min | ||
+ | |||
+ | ==== After power failure | ||
+ | |||
+ | * reset chiller complaining about closed outlet | ||
+ | * (on-off HT breaker at the back of power supply unit) | ||
+ | * press (longer) " | ||
+ | * (Lenses on - front of the microscope) | ||
+ | * (EDX detector controller on-off, while SIP is not running (ACD on)) | ||
+ | * watch vacuum | ||
+ | * (HT conditioning: | ||
+ | * HT -> Auto HT & Emission -> Start up | ||
+ | |||
+ | ==== Normal vacuum levels ==== | ||
+ | |||
+ | | Column/PIG1 | ~25 uA | SIP starts at ? uA | | ||
+ | | Specimen/ | ||
+ | | Camera/PIG3 | ~25 uA | ||
+ | | RT1/PIG5 | ~80 uA | ||
+ | | RT2/PIG6| ~25 uA | ||
+ | | 20l SIP | <1 uA | | ||
+ | | 60l SIP | ~1 uA | | ||
+ | | Column SIP | 200 uA | 800 uA with specimen without LN2 in ACD | | ||