Line 21: Line 21:
 ==== Alignment ==== ==== Alignment ====
  
-  * GUNA(F3)center with DEF @ Spot1 and SHIFT Spot5, or Anode wobler(F5)+  * GUNA(F3) Spot1 center with SHIFT <-> BRIGHT TILT Spot5 center with SHIFT 
 +  * GUNA(F3) and Anode wobler(F5) make symetric with DEF
   * COND STIG: make round with DEF   * COND STIG: make round with DEF
-  * Tilt and shift purity 
   * STD FOCUS adjust Z (minimal contrast, no hallo, or no motion with IMAGE WOBB)   * STD FOCUS adjust Z (minimal contrast, no hallo, or no motion with IMAGE WOBB)
 +  * Tilt and shift purity (at correct focus, and diffraction settings!)
   * OBJ wobler (F4) make feature stationary with DEF(BRIGHT TILT)   * OBJ wobler (F4) make feature stationary with DEF(BRIGHT TILT)
   * make fft round with DEF(OBJ STIG)   * make fft round with DEF(OBJ STIG)
-  * Filter->Slit->IN center circle with Isochromacity +  * SPECTRUM->move ZLP to center using FL lens 
-  * center diffraction win DEF(PLA)+  * Filter->Slit->IN move slit to center using position 
 +  * back to IMAGE, center circle with Isochromacity 
 +  * center diffraction with DEF(PLA) 
 +  * focus diffraction with DIFF FOCUS at Alpha 3, camera length 200cm, condensor fully CW
  
  
 ==== After bakeout ==== ==== After bakeout ====
  
 +  * remove mechanical safeties from EDX detector valve and beam valve
   * Cycle BEAM VALVE with HT OFF   * Cycle BEAM VALVE with HT OFF
   * Open EDX detector valve with BEAM VALVE closed   * Open EDX detector valve with BEAM VALVE closed
 +  * Conditioning knob -> COND
   * HT->HT Conditioning->Auto Procedure->Start (it takes ~60 min)   * HT->HT Conditioning->Auto Procedure->Start (it takes ~60 min)
 +  * Conditioning knob -> OPERATE
  
 ==== Beam shower ==== ==== Beam shower ====
  
-  * ?+  * Camera length 150cm 
 +  * CLA out, SAA #4, HCA #4 
 +  * CL1 0x8000, CL2 0x6000 (load STEMBeamShower.flc in FLC Panel) 
 +  * wait 10-15 min 
 + 
 +==== After power failure  ==== 
 + 
 +  * reset chiller complaining about closed outlet 
 +  * (on-off HT breaker at the back of power supply unit) 
 +  * press (longer) "I" on the front of the microscope 
 +  * (Lenses on - front of the microscope) 
 +  * (EDX detector controller on-off, while SIP is not running (ACD on)) 
 +  * watch vacuum 
 +  * (HT conditioning: switch knob to Cond, Auto procedure start, switch knob back to Operate) 
 +  * HT -> Auto HT & Emission -> Start up 
 + 
 +==== Normal vacuum levels ==== 
 + 
 +| Column/PIG1 | ~25 uA   | SIP starts at uA | 
 +| Specimen/PIG4 | ~25 uA | ~250 uA  without holder, pump LED goes green at 28 uA | 
 +| Camera/PIG3 | ~25 uA     | 
 +| RT1/PIG5 | ~80 uA     | 
 +| RT2/PIG6| ~25 uA    |  
 +| 20l SIP | <1 uA | 
 +| 60l SIP | ~1 uA | 
 +| Column SIP | 200 uA | 800 uA with specimen without LN2 in ACD | 
  
  