Line 51: Line 51:
   * wait 10-15 min   * wait 10-15 min
  
-==== After (brief) power failure  ====+==== After power failure  ====
  
   * reset chiller complaining about closed outlet   * reset chiller complaining about closed outlet
 +  * (on-off HT breaker at the back of power supply unit)
   * press (longer) "I" on the front of the microscope   * press (longer) "I" on the front of the microscope
 +  * (Lenses on - front of the microscope)
 +  * (EDX detector controller on-off, while SIP is not running (ACD on))
   * watch vacuum   * watch vacuum
 +  * (HT conditioning: switch knob to Cond, Auto procedure start, switch knob back to Operate)
   * HT -> Auto HT & Emission -> Start up   * HT -> Auto HT & Emission -> Start up
  