This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision | ||
tem:procedures [2016/02/09 10:05] admin [Alignment] |
tem:procedures [2021/10/25 08:53] (current) |
||
---|---|---|---|
Line 46: | Line 46: | ||
==== Beam shower ==== | ==== Beam shower ==== | ||
- | * ? | + | * Camera length 150cm |
+ | * CLA out, SAA #4, HCA #4 | ||
+ | * CL1 0x8000, CL2 0x6000 (load STEMBeamShower.flc in FLC Panel) | ||
+ | * wait 10-15 min | ||
- | ==== After (brief) | + | ==== After power failure |
* reset chiller complaining about closed outlet | * reset chiller complaining about closed outlet | ||
+ | * (on-off HT breaker at the back of power supply unit) | ||
* press (longer) " | * press (longer) " | ||
+ | * (Lenses on - front of the microscope) | ||
+ | * (EDX detector controller on-off, while SIP is not running (ACD on)) | ||
* watch vacuum | * watch vacuum | ||
+ | * (HT conditioning: | ||
* HT -> Auto HT & Emission -> Start up | * HT -> Auto HT & Emission -> Start up | ||